Appendix
83
ELECTROSURGICAL UNIT PSD-30
Item Specification
Safety
functions
Self-check
monitor
The function detects whether the safety monitor
functions mentioned above are working correctly or
not. It activates the following functions if
abnomalities are detected:
All indicators on the front panel will flash.
An alarm will sound.
Change of setting will be prevented.
Output will be prevented.
Feedback ratio
monitor
The irregularities are detected by calculating the
ratio of delivered current (through either A-cord) to
return current (through patient plate and S-cord). If
the irregularities are detected, the following functions
are triggered:
High frequency output warning indicator will light.
If the above condition continues for a while, an
alarm will sound and high frequency output will be
prevented.
When using an endoscope without a S-cord, high
frequency output warning indicator will light, an
alarm will sound and high frequency output will be
prevented.
Additional
functions
Stand-by mode When output is not performed for a long period and
the stand-by mode is selected, the following
operations are prevented:
High frequency output.
Change of setting classification (electromedical
equipment)
Adjusting the
volume
The output volume can be adjusted in the stand-by
mode.
PROGRAM
mode
Call the setting and CUT/COAG. mode stored in the
memory.