Important Information — Please Read Before Use
11
UCR INSTRUCTION MANUAL
WARNING
Be sure that this instrument is not used adjacent to or stacked with other equipment
(other than the components of this instrument or system) to avoid electromagnetic
interference.
Electromagnetic interference may occur on this instrument near equipment marked
with the following symbol or other portable and mobile RF (Radio Frequency)
communications equipment such as cellular phones. If electromagnetic
interference occurs, mitigation measures may be necessary, such as reorienting or
relocating this instrument, or shielding the location.
CAUTION
Do not use a pointed or hard object to press the buttons on the front panel. This
may damage the instrument.
To prevent equipment damage, do not use it in a dusty environment.
Use in combination with the upper/lower gastrointestinal
endoscope
WARNING
Over-insufflating the lumen may cause patient pain, injury, bleeding, gas embolism
and/or perforation.
Anytime you observe an irregularity in the endoscopic CO
2
regulation unit,
immediately stop using the endoscopic CO
2
regulation unit and withdraw the
endoscope from the patient slowly as described in the endoscope’s instruction
manual.
•The CO
2
gas is normally emitted through the small hole in the endoscope’s
air/water valve. Persons in the operating room may be affected if the CO
2
concentration in the operating room increases. Be sure to ensure ventilation of the
room.
During use, always stop the device like light source from supplying air. If “Stop” is
not selected, a mixture of air and CO
2
may be supplied into the patient body.
To prevent the water in the water feed tank from flowing into this unit through the
gas tube, install this unit in a position as high as possible compared to the water
feed tank.