5.5 After use
51
UCR INSTRUCTION MANUAL
Ch.5
Disconnecting the medical gas pipeline
1 Disconnect the hose from the CO
2
gas connector of the medical gas pipeline.
2 Disconnect the water container-side connector of the gas tube from the gas tube
connector on the water container (MAJ-902).
3 Disconnect the Gas tube (MAJ-1741), Low flow gas tube (MAJ-1742), or Extra low
flow gas tube (MAJ-1816) from the CO
2
gas outlet of this instrument.
4 Press the start/stop switch of the endoscopic CO
2
regulation unit to start the gas
insufflation for removing CO
2
gas remaining inside the instrument.
5 After the endoscopic CO
2
regulation unit has entered the stop mode, turn it OFF.
6 Disconnect the power cord plug from the hospital grade AC outlet (wall mains outlet).