4.2 Inspection workflow
UCR INSTRUCTION MANUAL
Ch.4
See the inspection workflow in Figure 4.1 below. Follow each step of the workflow for inspection of the
endoscopic CO
2
regulation unit before use.
4.2 Inspection workflow
1
Inspect the ancillary equipment and optional items.
Cylinder hoseSection 4.3 on page 37
Medical gas pipeline adapterSection 4.4 on page 37
Gas tube
*1
Section 4.5 on page 38
*1 When the lens cleaning sheath is used, the gas tube is not used so it therefore does not
need inspection.
2
Confirm that the endoscopic CO
2
regulation unit turns ON
normally.
on page 38
3
Confirm that a gas cylinder or the medical gas pipeline adapter is
connected.
on page 39
4
Confirm that CO
2
gas is delivered from the CO
2
gas outlet.
on page 40
5
Confirm that CO
2
gas is delivered from the gas tube.
*1
on page 41
6
Inspect the gas and water feeding function.
*2
on page 42
*2 When the lens cleaning sheath is used, the gas and water feeding functions should be
inspected by following the instruction manual for the lens cleaning sheath.
7
Turn OFF the endoscopic CO
2
regulation unit.
on page 43