38
4.5 Inspection of the Gas tube (MAJ-1741), Low flow gas tube (MAJ-1742), and Extra low flow
gas tube (MAJ-1816)
UCR INSTRUCTION MANUAL
Ch.4
WARNING
If the gas tube is damaged, replace it with a new tube.
Water drops remaining on or inside the tube could cause damage to internal
sensors (e.g., short circuit) or cause an electric shock. Dry the tubes thoroughly
before use.
Make sure the inside of the tube is clean. Using a dirty tube can cause infections in
the patient. If it is dirty, dispose of it according to the instructions in Chapter 6,
“Reprocessing, Storage, Disposal, and Transportation” and replace it with a new
one.
Inspection of the power supply
CAUTION
If the power indicator does not light up, the instrument may be damaged.
Immediately turn the endoscopic CO
2
regulation unit OFF, disconnect the power
plug from the hospital grade receptacle (wall mains outlet) and contact Olympus.
4.5 Inspection of the Gas tube (MAJ-1741), Low flow
gas tube (MAJ-1742), and Extra low flow gas
tube (MAJ-1816)
1 Check the gas tube ID on the tag and confirm that the gas tube matches the selected
flow rate.
2 Inspect the tube and connector for scratches, cracks, or other damage. Discard and
replace any damaged equipment.
3 Confirm that the tube and the connector are dry.
4.6 Inspection of the endoscopic CO
2
regulation
unit
1 Press the power switch to turn the endoscopic CO
2
regulation unit ON.
2 Confirm that the power indicator lights up.