6.1 Reprocessing
55
UCR INSTRUCTION MANUAL
Ch.6
CAUTION
Do not reprocessing the AC mains power inlet. Reprocessing them can deform or
corrode the contacts, which could damage the endoscopic CO
2
regulation unit.
Do not soak in water and autoclave, the endoscopic CO
2
regulation unit and
accessories. These methods will damage them.
Do not wipe the external surface with hard or abrasive wiping material. The surface
will be scratched.
When patient debris enters a hole or gap of the endoscopic CO
2
regulation unit,
contact Olympus without disinfecting it. If you try to disinfect by force, the
disinfectant solution will get inside the endoscopic CO
2
regulation unit causing fire
and malfunction of the endoscopic CO
2
regulation unit.