App.
70
Specifications
UCR INSTRUCTION MANUAL
Supply pressure indications • Five steps by LEDs.
5: above 4.5 MPa
4: 3.3 – 4.5 MPa
3: 2.3 – 3.3 MPa
2: 1.3 – 2.3 MPa
1: 0.3 – 1.3 MPa
• When the cylinder pressure drops below 0.3 MPa, the cylinder
pressure indicator will light red.
Start/stop • When the start/stop switch is pressed, the gas supply indicator
illuminates and the unit starts supplying the gas.
• When the start/stop switch is pressed again, the indicator goes
out and the unit stops supplying the gas.
Gas feeding pressure Maximum pressure
feed
45 kPa
Timer Timer setting:
After the set time elapses, the gas supply stops.
Classification
(electromedical
equipment)
Type of protection
against electric
shock
Class I (3-pin power cord)
Degree of protection
against electrical
shock of applied part
Type BF applied part:
Gas tube (MAJ-1741)
Low flow gas tube (MAJ-1742)
Extra low flow gas tube (MAJ-1816)
EMC Applied standard IEC 60601-1-2: 2001
IEC 60601-1-2: 2007
IEC 60601-1-2: 2014
• This instrument complies with the EMC standard for medical
electrical equipment, edition 4 (IEC 60601-1-2: 2014).
When connecting to an instrument that complies with a previous
edition of the EMC standard for medical electrical equipment
edition, the EMC characteristics could be vulnerable.
• CISPR 11 of emission:
Group 1, Class B
Year of manufacture The last digit of the year of manufacture is the second digit of the
serial number. In this example, the year is 2006.
Ex. 7612345 (serial number)
Medical Devices Directive
This device complies with the requirements of Directive 93/42/EEC
connecting medical devices.
Classification: Class II a
Item Specification