EasyManua.ls Logo

Olympus UCR - Page 82

Olympus UCR
86 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
App.
76
EMC information
UCR INSTRUCTION MANUAL
NOTE
At 80 MHz and 800 MHz, the higher frequency range applies.
These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and people.
Electromagnetic interference may occur in the vicinity of high-frequency
electrosurgical equipment and/or other equipment marked with the following
symbol:
Immunity
test
IEC 60601-1-2
(2014)
test level
IEC 60601-1-2
(2007, 2001)
test level
Compliance
level
IEC 60601-1-2 (2007, 2001)
Electromagnetic environment —
Guidance
Conducted RF
IEC 61000-4-6
3V
(150 kHz –
80 MHz)
3V (V
1
)
(150 kHz –
80 MHz)
Same as left Recommended separation distance
Where “P” is the maximum output power
rating of the transmitter in watts [W]
according to the transmitter manufacturer
and “d” is the recommended separation
distance in meters [m].
6V
(ISM band of
150 kHz –
80 MHz)
Same as left
ISM (industry, science, and medical care) band of 6.765 MHz – 6.795 MHz, 13.553 MHz –
13.567 MHz, 26.957 MHz – 27.283 MHz, and 40.66 MHz – 40.70 MHz between 0.15 MHz and
80 MHz
Radiated RF
IEC 61000-4-3
3V/m (80 MHz –
2.7 GHz)
3V/m (E
1
)
(80 MHz –
2.5 GHz)
Same as left Recommended separation distance
80 MHz – 800 MHz
800 MHz – 2.5 GHz
Where “P” is the maximum output power
rating of the transmitter in watts [W]
according to the transmitter manufacturer
and “d” is the recommended separation
distance in meters [m].
Proximity
magnetic field
from RF
communication
equipment
IEC 61000-4-3
Refer to the table
of the next page.
Same as left
d
3,5
V
1
-------
P=
d
3,5
E
1
-------
P=
d
7
E
1
------
P=

Table of Contents

Related product manuals