Chapter 10 Troubleshooting
73
VISERA RHINO-LARYNGO VIDEOSCOPE ENF-V2
10.1 Troubleshooting guide
The following table shows possible causes of and countermeasures against
troubles that may occur due to equipment setting errors or deterioration of
consumables.
Troubles or failures due to other causes than those listed below should be
serviced. As repair performed by persons who are not qualified by Olympus
could cause patient or operator injury and/or equipment damage, be sure to
contact Olympus for repair according to Section 10.3, “Returning the endoscope
for repair” on page 75.
Image quality or brightness
Other
Irregularity
description
Possible cause Solution
There is no video
image.
Not all the power
switches are ON.
Turn ON all the power switches.
An image is not clear. The distal end is dirty. Clean the distal end with a cotton
swab moistened with 70% ethyl or
isopropyl alcohol.
An image is excessively
dark or bright.
The light source is not
set properly.
Adjust the light source as
described in its instruction
manual.
Irregularity
description
Possible cause Solution
The remote switch does
not work.
Wrong remote switch is
operated.
Operate the correct remote
switch.
The remote switch
function has been set
incorrectly.
Set the remote switch function
correctly.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Olympus VISERA ENF-V2 and is the answer not in the manual?

Olympus VISERA ENF-V2 Specifications

General IconGeneral
BrandOlympus
ModelVISERA ENF-V2
CategoryMedical Equipment
LanguageEnglish

Related product manuals