80
Appendix
VISERA RHINO-LARYNGO VIDEOSCOPE ENF-V2
Guidance and manufacturer’s declaration —
Electromagnetic immunity
This model is intended for use by medical personnel in hospital
environments and for use in the electromagnetic environment specified
below. The customer or the user of this model should assure that it is used
in such an environment.
Immunity test
IEC 60601-1-2
(2014)
test level
IEC 60601-1-2
(2007, 2001)
test level
Compliance
level
IEC 60601-1-2 (2007, 2001)
Electromagnetic environment —
Guidance
Electrostatic
discharge
(ESD)
IEC 61000-4-2
Contact:
8kV
Air:
2, 4, 8,
15 kV
Contact:
2, 4, 6kV
Air:
2, 4, 8kV
Same as left Floors should be made of wood, concrete,
or ceramic tile that hardly produces static.
If floors are covered with synthetic material
that tends to produce static, the relative
humidity should be at least 30%.
Electrical fast
transient/burst
IEC 61000-4-4
2kV
for power supply
lines
1kV
for input/output
lines
2kV
for power supply
lines
1kV
for input/output
lines
Same as left Mains power quality should be that of a
typical commercial (original condition
feeding the facilities) or hospital
environment.
Surge
IEC 61000-4-5
Differential mode:
0.5, 1kV
Common mode:
0.5, 1, 2kV
for signal input/
output lines:
2kV
Differential
mode:
0.5, 1kV
Common mode:
0.5, 1, 2kV
Same as left Mains power quality should be that of a
typical commercial or hospital
environment.