EasyManuals Logo

Omega DP20 User Manual

Omega DP20
30 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #11 background imageLoading...
Page #11 background image
11
DP20 Economical Multisignal meter Instruction Manual
1.13 Measuring with Pt and Ni probes
Accepted Pt and Ni probes
The instrument accepts connecon of Pt100, Pt500 and
Pt1000 temperature probes, and also Ni100, Ni200 and
Ni1000 temperature probes. Temperature ranges for
each type of probe are indicated on Table 11’ below.
Pt100 with 2 and 3 wires
The instrument accepts connecon for 2 and 3 wire Pt100 probes.
For 3 wire Pt100 probes, congure internal jumper T’ at posion
5-6 (see secon 1.9). For 2 wire Pt100 probes, congure internal
jumper ‘T’ at posion 4-5. Compensaon of the wire resistance for 2
wire probes can be manually congured with the parameter Oset
reading(‘oFFS’) (see secon 1.21) which allows to congure a xed
number of counts to be added to the reading.
Resoluon and units
The temperature resoluon using Pt and Ni temperature probes is
congurable to 1º or 0.1º. Reading can be congured in ºC (degrees
Celsius) or ºF (degrees Fahrenheit).
Sensor break detecon
In case of sensor break, the instrument will show h.ovror h.udr
(see Table 18) depending on the broken cable.
Alpha temperature coecient
The instrument accepts Pt probes (Pt100, Pt500, Pt1000) with
Alphaparameter of 0.0385and ‘0.0390(see secon 1.28.7). This
parameter is associated to the specic model of probe installed.
Sensor
Jumper ‘S’
(see secon 1.9)
Jumper ‘T’
(see secon 1.9)
Range in ºC
(in ºF)
Total error
Connecon
(
terminals)
Current at
sensor
Pt100 3 wires F H J 5-6
-200 / 700 ºC
(-328 / 1292 ºF)
<1º
2 (Pt+)
4 (Pt -)
5 (sense)
< 900 uA
Pt100 2 wires F H
4-5
-200 / 700 ºC
(-328 / 1292 ºF)
2 (Pt+, Ni+)
4 (Pt -, Ni-)
Pt500 F
-150 / 630 ºC
(-238 / 1166 ºF)
< 90 uA
Pt1000 F
-190 / 630 ºC
(-310 / 1166 ºF)
< 90 uA
Ni100 F H
-60 / 180 ºC
(-76 / 356 ºF)
< 900 uA
Ni200 F H
-60 / 120 ºC
(-76 / 248 ºF)
< 900 uA
Ni1000 F
-60 / 180 ºC
(-76 / 356 ºF)
< 90 uA
Table 11 - Ranges of temperature for Pt and Ni probes
Terminal 5 ‘mulfuncon’ - ‘Pt100 3 wire’ or ‘External control’
To congure for 3 wire Pt100 probes, set internal jumper T at
posion 5-6 (see secon 1.9). Terminal 5 will be assigned to the
connecon of the third wire of the Pt100 probe, for automac wire
resistance compensaon purposes, up to 30 R.
To congure for 2 wire Pt100 probes, set internal jumper T at
posion 4-5 (see secon 1.9). Terminal 5 will be congured as ‘EK’
external control funcon. See secon 1.9 for a list of available
funcons.
Start-up, connecons and jumpers
For instrument start-up follow the steps listed at secon 1.7. Signal
connecons are indicated at secon 1.5. Locaon for internal
jumpers is indicated at secon 1.9.
See below a list of typical connecons :
Pt+, Ni+
Pt-, Ni-
12345
other probes 2 wires Pt100 3 wires
Pt-
Sense
Pt+
12345

Other manuals for Omega DP20