• Accepted NTC probes
The NTC probe is a temperature variable resistor
with a temperature - resistance curve dened by two
parameters called ‘R25’ and ‘beta’. By default, the
instrument is congured for a standard NTC with ‘R25’
of 10K and ‘beta’ of 3500. The measured temperature ranges from
-60 ºC to 150ºC.
Dierent NTC probes can be used, by conguring the instrument for
dierent values of ‘R25’ and ‘beta’. The instrument measures
resistance from 100 R up to 1 MOhm. Use ‘Table 12’ to write down
values for your NTC.
• Parameters ‘R
25
’ and ‘Beta’
The instrument accepts values of ‘beta’ between 2500 and 5500,
and values of R25 between 1.0 K and 200.0 K
• Resoluon and units
The temperature resoluon using NTC temperature probes is
congurable to 1º or 0.1º. Reading can be congured in ºC (degrees
Celsius) or ºF (degrees Fahrenheit).
Range
of measure
NTC probe
Temp. R25 Beta
100 R
1 MOhm
Table 12 - Data from NTC datasheet
• Sensor break detecon
In case of sensor break, the instrument will show ‘h.ovr’ or ‘h.udr’
(see Table 18) depending on the broken cable.
• Terminal 5 ‘mulfuncon’ - External control
Terminal 5 remains congured as ‘EK’ external control funcon. See
secon 1.9 for a list of available funcons.
• Start-up, connecons and jumpers
For instrument start-up follow the steps listed at secon 1.7. Signal
connecons are indicated at secon 1.5. Locaon for internal
jumpers is indicated at secon 1.9.
See below connecons for NTC probe :
1.14 Measuring with NTC probes
NTC ‘R
25
’
(congurable)*
Jumper ‘S’
(see secon 1.9)
Jumper ‘T’
(see secon 1.9)
Range
of measure
Accuracy
(% of reading)
Connecon
(
terminals)
Beta
(congurable)
10K F K 4-5
from -60
º
C
to 150
º
C
<1.5% of
reading
2 (NTC +)
4 (NTC -)
3500
Table 13 - *‘Beta’ is congurable from 2000 to 5500. ‘R25’ is congurable from 1.0 K up to 200.0 K.
ntc +
ntc -
12345
1.15 Measuring with PTC probes
• Accepted PTC probes
The instrument accepts direct connecon of PTC
probes. Accepted PTC probes are listed at ‘Table 14’.
• Resoluon and units
The temperature resoluon using PTC temperature probes is
congurable to 1º or 0.1º. Reading can be congured in ºC (degrees
Celsius) or ºF (degrees Fahrenheit).
Family Sensor
Jumper ‘S’
(see secon 1.9)
Jumper ‘T’
(see secon 1.9)
Range in ºC
(in ºF)
Total error
KTY 121
KTY81-121
KTY82-121
F
4-5
-55 / 150 ºC
(-67 / 302 ºF)
<1ºKTY 210
KTY81-210
KTY82-210
FHK
KTY 220
KTY81-220
KTY82-220
FHK
Table 14 - Ranges of temperature for PTC probes
• Sensor break detecon
In case of sensor break, the instrument will show ‘h.ovr’ or ‘h.udr’
(see Table 18) depending on the broken cable.
• Terminal 5 ‘mulfuncon’ - External control
Terminal 5 remains congured as ‘EK’ external control funcon. See
secon 1.9 for a list of available funcons.
ptc +
ptc -
12345
• Start-up, connecons and jumpers
For instrument start-up follow the steps listed at secon
1.7. Signal connecons are indicated at secon 1.5.
Locaon for internal jumpers is indicated at secon 1.9.
See below connecons for PTC probe :