EasyManuals Logo

Omron FQ2-S Series User Manual

Omron FQ2-S Series
444 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #291 background imageLoading...
Page #291 background image
Command Control
FQ2-S/CH User’s Manual
for Communications Settings
289
5
Appendices
GETDICFIGPARAM or GDFP
Acquire Dictionary Data Cutout Region Parameters
This command acquires the specified dictionary data cutout region parameters.
<Command Format>
<Response Format>
When the Command Is Processed Normally
When the Command Is Not Processed Normally
<Parameter Descriptions>
Dictionary data number Specifies the dictionary data number. (0 to 31)
Region number With the FQ2, specify 0.
Figure number With the FQ2, specify 0.
Parameter 1 to parame-
ter 5
The items of information for the dictionary data cutout region are split up and returned in
parameter 1 through parameter 5.
*
GE F I GPARAM
CR
TD CI
Figure number
(fixed value of 0)
Dictionary data number
(2 digits max.)
Command
Space
(0x20)
Space
(0x20)
Space
(0x20)
Region number
(fixed value of 0)
Space
(0x20)
Space
(0x20)
Dictionary data number
(2 digits max.)
Command
or
DGPF
Region number (fixed value of 0)
Figure number
(fixed value of 0)
CR
Delimiter
Delimiter
Space
(0x20)
CR
OK
Delimiter
Comma (fixed separator character)
Comma (fixed separator character)
Parameter 1 Parameter 2 ,
CR
... Parameter 5,
CR
ER
Delimiter
FQ2-S_CH_comm.book 289 ペー 01年626 木日 前1時4