EasyManua.ls Logo

Omron RX - Page 120

Omron RX
414 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
4-15
4-2 Function Mode
4
Functions
Adjusting the O2-L Terminal
* Below is the ratio from -10 to 10 V.
-10 to 0 V: -100% to 0%
0 to 10 V: 0% to 100%
For example, to input voltage ranging from -5 to 5 V on the
O2-L terminal, set A113 to -50%, and A114 to 50%.
O, O2, OI Sampling
You can set the built-in filter applied to frequency setting signals of the external voltage/current
input.
Helps remove noise in the frequency setting circuit.
Set a larger data value if stable operation cannot be secured because of noise.
Note that the larger the data value is, the slower the response time. This parameter specifies a filter
time constant for a set value of 1 to 30 (× 2 ms).
When "31." is selected, a filter time constant of 500 ms and a hysteresis of ±0.1 Hz are set. (Factory
default)
Parameter No. Function name Data
Default
setting
Unit
A111 O2 start frequency
-400. to 400.
(Set a start frequency.)
0.00
Hz
A112 O2 end frequency
-400. to 400.
(Set an end frequency.)
0.00
A113 O2 start ratio
-100. to 100.:
(Set a start ratio relative to an external
frequency reference of -10 to 10 V.
*
)
-100.
%
A114 O2 end ratio
-100. to 100.:
(Set an end ratio relative to an external
frequency reference of -10 to 10 V.
*
)
100.
Related functions A003, A203, A303, A081, A082
Forward
max. frequency
A112
Analog input
(02)
(Example 3)
(-10 V)
-100%
A113
A111
A114
100%
(+10 V)
Reverse
max. frequency
Parameter No. Function name Data Default setting Unit
A016 O, O2, OI sampling 1. to 30./31. 31. Time
Related functions A011 to A016, C001 to C008

Table of Contents

Related product manuals