OMS LINEA ESSE S.T.01/3S ED.0 REV.14 03/2017
16
3. COMMAND - CONTROL - WARNING
SAFETY SYSTEMS
Care should be taken when driving the patient chair to minimize the risk of collision during the movement,
Obstruction in the patient chair’s line of movement activate safety switches that stop the motorized
movements as well as cuspidor when is out.
PERICOLO
CHAIR MOON The backrest has also a safety microswitch located on backrest
itself, (inside the seat, on the lower part of the backrest) as the sensor before it while
the chair or backrest is driving down, the movement get stop and a sounds beep (bip-
bip) inform that the safely backrest chair is activated
DANGER
In the version PLUS with a floor cuspidor, the bowl is above the patient chair and stops
upward chair movements, Move the bowl into home position to resume normal
operation.
Note:
To calibrate the backrest’s sensore from the Chair MOON:
Drive the chair or complete dental unit, into the upper position.
Use a small flat screew driver, inside the hold , and turn the trimmer , turn it clockwise to
decrease the sensibility or turn it anti clockwise to increase the senibility. To calibrate the
backrest’s sensore from the MOON Chair, follow the procedur.
Drive the chair or complete dental Unit, into the upper position.
Use a small flat screew driver, inside the hold , and turn the trimmer , (see the following
picure ), turn it clockwise to decrease the sensibility or turn it anti clockwise to increase
the senibility.
Drive towards down the chiar and by touching the backrest the saftly should activated.