INSTALLATION
VERSION WITH CUSPISOR CONNECTED TO THE CHAIR.
1. Remove the equipment from its packaging;
2. Install the chair, fix the seat, headrest and armrest if available. Fix the chair to the floor with the two
expansion bolts supplied;
3. Lift the cuspidor, fixing it to the chair with the supplied screws, it may be better do not remove the
packaging until the unit is fixed to the chair.
4. Open the cover of the cuspidor, and place the instrument arm with small rotating movements.
5. Connect the doctor console and cuspidor. Make sure that "Panduit" connectors are correct installed
to the numbers of the wires.
6. Remove the packaging from the doctor console and install accessories.
7. Connect transformer and dental unit to the junction box, make sure that the numbers of wires are
correctly connected. Make sure that “Panduit" connector correctly.