6. MAINTENANCE AND KEEP UP
To guarantee the dental unit’s proper operation, the unit must be checked and serviced by a qualified OMS
services technician according to the maintenance schedule that has been set for your dental units.
In the annual maintenance, “CARE KIT” the service technician replaces all parts specified by the Care Kit. These
include, but are not limited to, parts in contact with air, water and suction system. In additional the service
technician checks and services all dental units parts suspect to wear and tear in normal use. There includes
part in the cuspidor, foot control, instrument console, arms, patient chair, assistant arm and operation light.
Also, the mechanical stability and electrical safety inspection is performed.
The default maintenance interval is 365 days. The message “Care Kit” will remain you the annual maintenance
well in advance.
The activities are classified in:
Cleaning/disinfecting,
adjustments,
scheduled maintenance,
extraordinary maintenance.
CLEANING AND DISINFECTING
For hygiene (and to avoid stains of corrosive agents) clean the device frequently.
For hygiene and cleaning, without risk of damage, O.M.S. recommends the use of products that contain:
Quaternary Ammonium
Phenolic compounds
Iodophors
That do not contain:
Alcohol
Hypochlorite
Soda
Organic Solvents