OMS LINEA ESSE S.T.01/3S ED.0 REV.14 03/2017
39
START UP AFTER IDLE TIME MORE THAN 4 WEEKS. (DENTAL UNIT WITHOUT
DECIVES)
Following the step illustrate before (See Point 5.1)
Replace the water filter of the unit.
In case of the dental units has been idle for more than 10 weeks, is strongly recommend perform the
maintenance CARE KIT.
This operation must be performed by an authorized technician.
CAUTION
The flushing without specific device for water treatment, is not an operation anti-
legionella.
START UP THE UNIT WITH DESINFECTION SYSTEM WEK OR WEK LIGHT,
AFTER IDLE TIME LESS THAN 4 WEEKS. (DENTAL UNIT WITH WATER SYSTEM
DEVICE TREATMENT)
CAUTION
Before to use the water system device read careful the instruction manual of METASYS WEK
or WEK LIGHT.
This document is destined for dental care professionals.
CAUTION
It is mandatory check once in a week, as well as after idle long period unit the amount of
peroxide. Using the stick 464441“ the amount of the peroxide can be test in just 45 seconds.
CAUTION
It is mandatory perform the maintenance kit of system Wek or Wek Light (Cod 463329 e
511448) once a year, to ensure a reliable functioning.
Switch ON the unit.
Check that not leaks are present where it is possible.
Check the control panel of the WEK, make sure that not alarm is shown, if any alarm is present read the
instruction manual of Metasys WEK system.
Check the expired date of the pouch of GREEN&CLEEN, if needed replace it.
Make a flushing with all instrument installed on the doctor console, flushing should be at least 1 minute for
each instrument. Do not forget to set up the WEK as intensive mode.