OMS LINEA ESSE S.T.01/3S ED.0 REV.14 03/2017
38
5. START UP THE UNIT AFTER LEFT IDLE TIME
Following the instruction about how to start up the unit after the device has left idle for extended period.
NOTE:
Following the instruction of the device installed. For further information please refer the
instruction manual of the manufacture’s device.
START UP AFTER IDLE TIME LESS THAN 4 WEEKS. (DENTAL UNIT WITHOUT
DECIVES)
Switch on the unit.
Check that not leaks are present where it is possible.
Make sure that the water tap is open as well as the instruments water taps. Opportunely closed before.
Make a flushing with all instrument installed on the doctor console, flushing should be at least 1 minute for
each instrument.
Note:
Check that all instruments tap are open on the bottom of doctor console.
Fill the cup at least 5 time.
Use the bowl rinsing for at least 1 minute.
CAUTION
The flushing without specific device for water treatment, is not an operation anti-
legionella.
CAUTION
We strongly recommend follow this procedure (Point 5.1) daily.