EasyManua.ls Logo

OMS LINEA ESSE - (Dental Unit with Water System Device Treatment); Start up the Unit with Suction Amalgama Separator, Idle for more than 4 Settimane

OMS LINEA ESSE
77 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
OMS LINEA ESSE S.T.01/3S ED.0 REV.14 03/2017
41
START UP THE UNIT WITH DESINFECTION SYSTEM WEK OR WEK LIGHT,
AFTER IDLE TIME MORE THAN 4 WEEKS. (DENTAL UNIT WITH WATER SYSTEM
DEVICE TREATMENT)
Following the step illustrate before (See Point 5.3)
Replace the water filter of the unit.
In case of the dental units has been idle for more than 10 weeks, is strongly recommend perform the
maintenance CARE KIT.
This operation must be performed by an authorized technician.
CAUTION
In case that the verification of the peroxide fail, repeat the process of “intensive
decontamination set up”. If the test fail again call an authorized technician O.M.S..
Is strongly recommend perform the Maintenance KIT METASYS WEK (Cod. 463182)
START UP THE UNIT WITH SUCTION AMALGAMA SEPARATOR, IDLE FOR
MORE THAN 4 SETTIMANE.
CAUTION
Before to use the suction system device read careful the instruction manual.
This document is destined for dental care professionals.
CAUTION
In case no suction system are installed, replace the suction filter.

Other manuals for OMS LINEA ESSE

Related product manuals