OMS LINEA ESSE S.T.01/3S ED.0 REV.14 03/2017
50
CHAIR
Note:
Daily cleaning and disinfection must be done with approved OMS products. DO NOT use
aggressive product for daily upholstery cleaning.
The deep cleaning of the upholstery, armrests and headrest shall carry out every 90 days, with products
suitable for the treatment and following this procedure.
1. Wipe the upholstery with EMULSIO.
2. Start to treat a small surface and check that there are no changes on the upholstery, then apply it on
the whole he upholstery.
3. Wipe the upholstery with a water to remove the EMUSION from the surface
4. Wait at least 90 days before repeat the treatment.
SUCTION HIGH VOLUME
To get the best performance of the suction system we strongly recommend, follow the maintenance
instructions below.
After each patient, we recommend use a clean hot water for a few seconds, to rinse the tubes. Handpiece
need to be clean: with a suitable detergent.
The suction handpiece supplied by O.M.S. can be sterilized at 134°C.
We strongly recommend to clean up the suction filter at the end of each working day, therefor the correct
suction device maintenance.
• Use the recommend cleaner product indicated by the manufacture suction devices
• Remove the cover filter holder and clean up the filter. Use a lubricant on the oring
• We strongly recommend change the filter once on the month, do so early if necessary.
• Never mix up the disinfection liquid with other disinfection products.