ALERT POTENTIAL ISSUE POSSIBLE SOLUTION
“Warn AOUT 2 Communications Fail” Analog output controller hardware reports
communications failure - Analog Out port 2
Contact ONICON. RMA needed
“Warn AOUT 2 Timeout” Analog output controller hardware reports
execution timeout - Analog Out port 2
Contact ONICON. RMA needed
“Warn Freq High Flow” Flow exceeds congured high ow threshold
(95% of max FreqOut ow)
• Conrm the frequency output full scale is equal to
or greater than the design max ow of the system.
• Go to Main Menu -> user conguration-> Input/
Ouput-> Frequency Output to adjust the full scale
of the meter
“Warn Low Internal Regulator” Internal power regulator below acceptable
threshold
Contact ONICON. RMA needed
“Warn Low Isolated Regulator” Analog I/O power regulator below acceptable
threshold
Contact ONICON. RMA needed
“Warn RTC Fail” Real time clock chip malfunction Contact ONICON. RMA needed
“Warn Elec 1 Flow Unstable” Minor reading irregularities detected - electrode 1 • Verify the meter is installed at the correct location
in the system and installed at the correct height
oset to ensure the sensor is in the correct portion
of the pipe.
• Verify system is properly bled and that no air is
trapped in the piping system where the sensor is
placed.
• Disconnect power from the device and check
Electrode 1 wiring connections at sensor and
transmitter ends of wire.
• Check continuity along each Electrode 1 wire to
verify wire connection between the sensor and
transmitter.
• Also, check Electrode 1 wiring for continuity
between other wires, ground, or either body. If
problem persists Contact ONICON support.
“Warn Elec 2 Flow Unstable” Minor reading irregularities detected - electrode 2