| Important information
(EMC). To avoid electromagnetic disturbances, the device may only be operated and serviced in accordance
with the user manual and using the components supplied by Optomed.
If the performance of Optomed Aurora is lost or degraded due to electromagnetic disturbances, it may
cause unexpected or adverse operation of this device. If a malfunction cannot be rectified, please contact
your local distributor or Optomed customer service.
CAUTION!
Use of this equipment adjacent to or stacked with other equipment should be avoided because it
could result in improper operation. If such use is necessary, this equipment and the other equipment
should be observed to verify that they are operating normally.
The device should not be operated in the vicinity of high-frequency surgical equipment.
CAUTION!
Use of accessories, transducers and cables other than those specified or provided by the manufacturer
of this equipment could result in increased electromagnetic emissions or decreased electromagnetic
immunity of this equipment and result in improper operation.
CAUTION!
Portable RF communications equipment (including peripherals such as antenna cables and external
antennas) should be used no closer than 30 cm (12 inches) to any part of the Optomed Aurora
including cables specified by the manufacturer. Otherwise, degradation of the performance of this
equipment could result.
Replacement cables may only be purchased at Optomed. The use of accessories, any converters or cables
which are not specified in this user manual or have not been purchased as spare parts from Optomed can
result in increased emissions or reduced immunity of the device.
MANUFACTURER’S DECLARATION – ELECTROMAGNETIC IMMUNITY
Optomed Aurora maintains basic safety and performance when used in the electromagnetic environment
specified below. The customer or the user of the device should ensure that it is used in such an environment.
IMMUNITY TEST IEC 60601 TEST LEVEL COMPLIANCE LEVEL ELECTROMAGNETIC ENVIRONMENT – GUIDANCE
Electrostatic
discharge (ESD)
IEC 61000-4-2
±8 kV contact
±15 kV air
±2 kV, ±4 kV,
±6 kV, ±8 kV
indirect
contact
±2 kV, ±4 kV,
±6 kV, ±8 kV
contact
±2 kV, ±4 kV, ±8
kV, ±15 kV air
Floors should be wood, concrete or ceramic
tile. If floors are covered with synthetic
material, the relative humidity should
be at least 30%.