Optomed Aurora 39
Important information |
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
Electrical fast
transient/burst
IEC 61000-4-4
±2 kV for power
supply lines
±2 kV for AC
power supply
Mains power quality should be that of
a typical commercial or hospital
environment.
Voltage dips,
short interruptions
and voltage
variations on
power supply
input lines
IEC 61000- 4-11
0% UT for 0.5
at 8 ф angles;
0% 1 cycle;
70% UT for 25
cycles;
0% for 5 sec
0 % UT for 0.5 at
8 ф angles;
0% 1 cycle;
70% UT for 25
cycles;
0% for 5 sec
Mains power quality should be that of a
typical commercial or hospital environment.
If the user of the Optomed Aurora requires
continued operation during power mains
interruptions, it is recommended that the
device be powered from an uninterruptible
power supply or a battery.
30 A/m 30 A/m
Power frequency magnetic fields
should be at levels characteristic
of a typical location in a typical
commercial or hospital environment.
Conducted RF
IEC 61000-4-6
3 Vrms
150 kHz to 80
MHz & 6 Vrms
ISM frequency
3 Vrms
150 kHz to 80
MHz & 6 Vrms
ISM frequency
Portable and mobile RF communications
equipment should be used no closer to
any part of Optomed Aurora, including
cables, than the recommended separation
distance calculated from the
equation applicable to the frequency
of the transmitter.
Recommended
separation distance
d = 1.2 √P
d = 1.2 √P 80 MHz to 800 MHz
d = 2.3 √P 800 MHz to 2.7 GHz,
where P is the maximum output
power rating of the transmitter in
watts (W) according to the transmitter
manufacturer and d is the recommended
separation distance in meters (m).
Field strengths from fixed RF transmitters,
as determined by an electromagnetic site
survey
A
, should be less than the compliance
level in each frequemcy range
B
.
Interference may occur in the
vicinity of equipment marked
with the following symbol:
Radiated RF
IEC 61000-4-3
3 V/m
80 MHz to 2.7 GHz
AM Modulation
& 9–28 V/m
385 MHz to
6.0 GHz
Pulse Modulation
3 V/m
80 MHz to 2.7 GHz
AM Modulation
& 9–28 V/m
385 MHz to
6.0 GHz
Pulse Modulation
Surge
IEC 61000-4-5
±1 kV line(s) to line(s)
±2 kV line(s) to earth
±1 kV dierential
mode
±2 kV common
mode
Mains power quality should be that
of a typical commercial or hospital
environment.