2.3 Measurement Specifications [LT models]
LT02
LT15
LT22
Temperature range (scalable)
-50...600 °C
-50...975 °C
Ambient temperature (head)
-20...130 °C
-20...180 °C
Storage temperature (head)
-40...130 °C
-40...180 °C
Spectral range
8...14 µm
Optical resolution
2:1
15:1
22:1
System accuracy
1) 2) 3)
±1,0 °C or ±1,0 %
Repeatability
1) 3)
±0,5 °C or ±0,5 %
Temperature coefficient
4)
±0,05 K/ K or ±0,05 %/ K
NETD
3) 5)
100 mK
50 mK
Response time (95 % signal)
150 ms
Warm-up time
10 min
Emissivity/ Gain
0,100...1,100 (adjustable via programming keys or software)
Transmissivity
0,100...1,100 (adjustable via programming keys or software)
Interface (optional)
USB (programmable adapter)
Signal processing
Average, peak hold, valley hold (adjustable via programming keys or software)
1)
at ambient temperature 235 °C; whichever is greater; at object temperatures >0 °C; response time=1s
2)
Accuracy for thermocouple output: ±2,5°C or ±1%
w ww . . co m
information@itm.com1.800.561.8187

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the optris CT and is the answer not in the manual?

optris CT Specifications

General IconGeneral
Brandoptris
ModelCT
CategoryThermometer
LanguageEnglish

Related product manuals