Electrical Installation 65-
You will find the calibration code on a label fixed on the head cable (near the electronics). Please
do not remove this label or make sure the code is noted anywhere. The code is needed if the
electronics has to be exchanged or in case of a necessary recalibration of the sensor.
After you have modified the head code a reset is necessary to activate the change.
[►6 Operating]
4.3.2 Sensing Head Cable
On all CT models (exception 3M, P3, P7) the sensing head cable can be shortened if necessary.
On the models 1M, 2M and CTfast the sensing head cable can be shortened by max. 3 m.
A shortening of the cable will cause an additional measuring error of about 0,1 K/ m.
The 3M models are only available with 3 m cable.
On the CT models LT02, LT02H and LT10H the head cable must not be moved during the
measurement.