-60 -
The USB socket on the side is only intended for setup and service and not for continuous use.
4.1.4 Power supply
Please use a stabilized power supply unit with an output voltage in the range of 8–36 VDC (CT 4M:
8-30 VDC) which can supply 100 mA. The ripple should be max. 200 mV.
Please do never connect a supply voltage to the analog outputs as this will destroy the
output!
The CT is not a 2-wire sensor!