Software CompactConnect/ CompactPlus Connect 91-
8.2.4 Saving of parameter settings
After power on of the CT sensor the flash mode is active. It means, changed parameter settings will be
saved in the CT-internal Flash-EEPROM and will be kept also after the sensor is switched off.
In case settings should be changed quite often or continuously the flash mode can be switched off by using
the following command:
Decimal:
112
HEX:
0x70
Data, Answer:
byte 1
Result:
0 Data will be written into the flash memory
1 Data will not be written into the flash memory
If the flash mode is deactivated, all settings will only be kept as long as the unit is powered. If the unit is
switched off and powered on again all previous settings are lost.
The command 0x71 will poll the current status.
You will find a detailed protocol and command description on the software package in the directory:
\Commands.