SAMPLE ANALYSIS - FLOWER
19. Remove cap from vial and pull at least 2ml of sample into syringe used
previously. Exact amount of sample isn’t important. Press forward arrow on
screen.
20. Place syringe on syringe filter at sample port by gently twisting it on about a
quarter turn. Press forward arrow on screen.
21. Slowly inject at least 1ml of sample into sample port. Exact amount isn’t
important. There will be some resistance when injecting a sample- this is
normal. Press forward arrow on screen.