SCALE CALIBRATION
10. Scale Calibration
1. If the scale has been used for more than a month or if the calibration weight does
not read between 9.997 and 10.003g then a scale calibration is required. Make
sure the scale is on a level surface and there is no significant air movement. Turn
scale on and wait until scale reads 0.000g.
2. Press and hold ON button for about 3 seconds until CAL is displayed on screen.
Press ON button once more.
3. After a few seconds, the display will flash 10.000. Add the first calibration
weight to the scale.