SCALE CALIBRATION
4. After a few seconds, the display will flash 20.000. Add the second calibration
weight to the scale along with the first. NOTE: The second weight is included in a
bag below the scale in the LightLab box.
5. After a few seconds, the display will read PASS. The calibration is now complete.
If the result reads FAIL, repeat all steps again.