SAMPLE ANALYISIS- INFUSED PRODUCT
18. Set valve to load. Press forward arrow on screen.
19. Remove cap from vial and pull at least 2ml of sample into syringe used
previously. Exact amount of sample isn’t important. Press forward arrow on
screen.
20. Place syringe on syringe filter at sample port by gently twisting it on about a
quarter turn. Press forward arrow on screen.