Orifice rim heater self-test failed/is not run
Orifice rim heater supply voltage is missing, incorrectly connected, or is outside
the specified range:
restore power supply to the orifice rim heater (replace fuse as necessary);
check supply voltage level and adjust as necessary;
check wiring of the orifice rim heater and correct as necessary.
Pipe housing (including built-in orifice rim heater) is not installed:
install the pipe housing.
Pipe housing without built-in orifice rim heater of another OTT Pluvio
2
S
is
accidentally installed:
install the pipe housing with built-in orifice rim heater.
No self-test is run:
check the “Self-test interval” (Modes 2, 3, 4) setting and correct as
necessary (using the OTT Pluvio² Operating Software);
force self-test by a “Power-Reset” once.
Orifice rim heater does not work/does not work as expected
Orifice rim heater supply voltage is missing, incorrectly connected, or is outside
the specified range:
restore power supply to the orifice rim heater (replace fuse as necessary);
check supply voltage level and adjust as necessary; make sure that output
power of power source is sufficient;
check orifice rim heater wiring and correct as necessary; make sure that
cable cross-section is sufficient;
check heater status (using the OTT Pluvio² Operating Software).
Pipe housing (including built-in orifice rim heater) is not installed:
install the pipe housing.
Pipe housing without built-in orifice rim heater of another OTT Pluvio² S is
accidentally installed:
install the pipe housing with built-in orifice rim heater.
Configuration of the orifice rim heater results in unexpected heater
performance:
check settings
mode of operation
orifice rim target temperature (modes #1, #2, #3, #4)
lower temperature limit (mode #2)
switch-on time (mode #3)
on-time (mode #3)
after-run time (mode #4)
and correct as necessary (using the OTT Pluvio² Operating Software).
OTT Pluvio² S system time is incorrect (e.g. after power supply failure):
correct system time (using the OTT Pluvio² Operating Software).