No/low precipitation output
No precipitation output
when USB interface is/was connected,
after emptying,
start-up/power failure:
wait for 5 minutes.
Collecting bucket overflow:
empty the collecting bucket.
Evaporation rate is very high (collecting bucket is empty and ambient
t
emperature is high):
add approx. 1 to 2 litres of water to the collecting bucket.
Apparently low output for NRT values:
wait for 5 minutes (delay caused by filter algorithm).
Increase in weight > approx. 12 mm within 6 seconds:
no precipitation output (filter algorithm detects foreign matter), since amount
of natural precipitation is exceeded; check precipitation gauge status:
Warning “+16” (using the OTT Pluvio² Operating Software).
Precipitation gauge probably failed:
perform Guided Accuracy Test (using the OTT Pluvio² Operating Software).
Precipitation gauge is not level:
check the bubble level indicator and correct adjustment as necessary.
Note: Check whether measured value used is suited for present application
(behaviour over time: Real-time output (RT), Non Real-time output (NRT), or
combination value (RT-NRT))! For a description of the individual measured values,
refer to Chapter 3.1. Use another measured value as necessary.
Incorrect precipitation output
Foreign matter (leafs, insects, …) have entered the collecting bucket (increase in
weight < approx. 12 mm within 6 seconds; any higher increase is detected as
foreign matter by the filter algorithm):
manually correct the amount of precipitation by deducting a respective
amount.
Base plate is incorrectly levelled:
Rotate the base plate so that the connector block of the orifice rim heater
is facing in South direction (North direction on the southern hemisphere)!
Refer to Chapter 5.6.
Anti-freeze agent without water was added (hygroscopic behaviour):
Add anti-freeze agent only as an aqueous solution (add 40 % of water)!
41