1.15 Electromagnetic immunity
In accordance with the IEC 60601-1-2 4
th
edition standard One is suitable for use in the electromagnetic environment
specified below.
The customer or user of the system must ensure that it is used in the said environment.
Immunity
test
IEC 60601-1-2
test level
Compliance
level
Electromagnetic environment
Electrostatic discharge
(ESD) IEC 61000-4-2
8 kV contact
2/4/8/15 kV air
IEC 60601-1-2
Test level
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30%
Radiated
electromagnetic field
IEC 61000-4-3
3 V/m
80 MHz to 2.7 GHz
IEC 60601-1-2
Test level
Portable and mobile RF communications
equipment should be used no closer to
any part of One including cable.
Minimum distance 30 cm
Electrical fast
transient/burst
IEC 61000-4-4
2 kV for power
supply lines
1 kV for input/output
lines > 3 m
IEC 60601-1-2
Test level
Mains power quality should be that of a
typical commercial or hospital
environment
Surge
IEC 61000-4-5
0.5/1 kV differential mode
0.5/1/2 kV common mode
IEC 60601-1-2
Test level
Mains power quality should be that of a
typical commercial or hospital
environment
Conducted
disturbances induced
by RF fields
IEC 61000-4-6
3 V
150 kHz to 80 MHz
6 V
ISM frequencies
IEC 60601-1-2
Test level
Portable and mobile RF communications
equipment should be used no closer to
any part of One, including cable.
Minimum distance 30 cm
Voltage dips, short
interruptions and
voltage variations on
power supply input
lines
IEC 61000-4-11
10 ms 0 % a 0°, 45°, 90°,
135°, 180°, 225°, 270°,
315°
20 ms 0% a
500 ms 70% a 0°
5 s 0%
IEC 60601-1-2
Test level
Mains power quality should be that of a
typical commercial or hospital
environment. If the user of the One
requires continued operation during
power mains interruptions, it is
recommended that the One be powered
from an uninterruptible power supply or
a battery
Power frequency
(50/60 Hz) magnetic
field
IEC 61000-4-8
30 A/m
IEC 60601-1-2
Test level
Power frequency magnetic fields should
be at levels characteristics of a typical
location in a typical commercial or
hospital environment