ONE User manual Page 8/44
1.11 Packaging and environment
Transport, storage and environment: the kit is supplied in protective packaging (protection against physical impacts
and antistatic packaging). It must be stored under the following conditions:
Ambient temperature: -10°C to +70°C / 14°F to 158°F
Relative humidity: <95% without condensation
Atmospheric pressure: 500hPa to 1060hPa
Operation: the kit has been designed for normal use under the following conditions:
Ambient temperature: +10°C to +35°C / 50°F to 95 °F
Relative humidity: 30% to 75%
Atmospheric pressure: 700hPa to 1060hPa
Equipment packaging for return to distributor: should a return to the distributor be necessary, make sure to
package the sensor and box kit in its original packaging after having cleaned it thoroughly.
Documentation loss: all kits are shipped with its documentation. Please contact your distributor for a replacement
manual if this documentation is lost.
1.12 Marking and labelling symbols
These symbols are used on the product labels and inform you about the compliance with standards and the technical
specifications of the component.
Direct current.
Type BF applied parts, IEC 60601-1
Insulation class II
Important information: follow the
instructions for use
The CE marking certifies that
this product complies with
European directive 93/42/EEC
and its revised versions..
Name and address of manufacturer
Product identification code
Serial number
Manufacturing date (year and month)
Sensor waterproofness standard,
EN/CEI 60529 regulation.
Only the part of the sensor putted
in month, complies with this
standard.
Storage condition: temperature
limitations.
Storage condition: relative humidity
limitations
Storage condition: Atmospheric
pressure limitations.