ONE – User manual Page 34/44
The following disinfectant agent has been validated with the sensor: CIDEX OPA solution (0.55% Ortho-
phthalaldehyde solution). The maximum soaking period is 24 hours.
5.2.2 Recommended decontaminating products outside North-America
The following disinfectants are compatible with the sensor and the part of the positioner or sensor connection cable
that is inserted in the mouth:
2% Sodium Hypochlorite (maximum immersion time of 24 hours)
Ethyl alcohol (maximum immersion time of 24 hours)
Quaternary ammonium
5.3 Maintenance
5.3.1 Computer data-protection
Your patient and image database must be backed-up to be able to recuperate them if needed (in case of hard disk
or computer problems). It’s recommended to do the back up once a week
Ask the advice of your IT specialist with regard to the backup system that is best suited to your computer
configuration (external or removable hard disk, CD-ROM or DVD writer, etc.). Test and store the copies in a safe
place. It’s recommended to do the system back up twice a month
NOTE: The device is not serviced or maintained while in use with the patient