Non clinical testing specifications
Spatial resolution
15pl/mm
Low-Contrast resolution
Minimum value measured: 0.2mm
Dynamic range
14bits (16384 grey levels).
Sensor saturated at 130ms with no absorbing material (no
object in the X-ray field - X-ray generator radiological
parameters 7mA/65kV)
Signal to noise ratio
Between 33 dB and 40dB
Contrast function
0,24 @ 10lp/mm