8. CLEANING AND DISINFECTING
This instrument requires routine cleaning, which includes removal of any soil or dirt from the
external surfaces. A soft cloth dampened lightly with water may be used.
In-between patients the Oxitone 1000M should be cleaned and low-level disinfected.
To clean, wipe the device surfaces with a soft cloth swab moistened in ethyl alcohol. Note that ethyl
alcohol is considered both a cleaning agent and a low level disinfecting agent
To low-level disinfect the device after each patient, follow these steps:
Remove the device from the patient.
After cleaning any visual debris from the surface of the device, again thoroughly wipe the
surface with an ethyl alcohol (70-85%) swab and a long Q-tip.
Allow the device to dry completely before returning it to operation.
Caution: The system is approved for IPX0. To avoid damage to device, be careful of
liquid spillage while cleaning.
Caution: To avoid damage to the surface of the instrument, take care not to scratch
the display while cleaning.
Caution: Turning off the device before cleaning.
Caution: Remove any substances or material that may interfere with the transmission
of light.
Caution: Clean and low level disinfect the device before applying it to a new patient.
Caution: Do not sterilize, autoclave, or immerse this device in liquid. Do not pour or
spray any liquid into the device.
Caution: Do not use caustic or abrasive cleaning agents, or any cleaning products
containing ammonium chloride, isopropyl alcohol, petroleum-based or acetone
solutions.
Caution: Do not use cleaning solution other than those recommended here, as
permanent damage could result.
Caution: Alcohols are volatile and flammable and must not be used near open
flames.