11. SPECIFICATIONS AND COMPLIANCE STANDARDS
Performance
Measurement Range
SpO
2
0%
-
100%
PR (pulse rate)
30
-
250 bpm
Accuracy
Arterial Oxygen Saturation, 70% to 100%
1
± 3%
Pulse Rate Accuracy
2
± 3 bpm
Resolution
Arterial Oxygen Saturation (SpO
2
)
1%
Pulse Rate
1 bpm
Skin temperature range
(d
irect mode)
29.0°C
-
43.0°C (84.2
°F
-
109.5°F)
Accuracy
±0.1°C
Interfering Substances
Refer to safety Inf
ormation, Warnings and Cautions
Product
Test time
30 seconds
Electrical
Battery Powered
Rechargeable lithium polymer
3.7VDC 370 mAh
Capacity
(opera
tion)
Approximately
24
hours after full charge
Number of spot checks on fully charged battery
Minimum 500
Battery charging time
Up to
2
h
our
s.
P
ower off
(min. 300 recharging
cycles)
Charger Isolation: class ll double isolation
5V AC/DC Adapter
AC P
ower for bench top charger
100
-
240V, 50
-
60 Hz, 10VA max
Measurement LED'S Wavelength and Output Power
Red
660nm @ 1.05 mW max. average
Infrared
940nm @ 0.95 mW max. average
Environmental
Operating temperature
32°F to 97°F (0°c to +36°C)
Operating hu
midity
5% to 95%, non
-
condensing
Pressure
700 to 1060 hPa
Operating altitude
-
378m to 3050m
(-1240 feet to 10000 feet)
Storage and transportation
Storage temperature
(
-
20°C to +60°C)
(
-
4°F to 140°F
)
Humidity
5% to 95% non
-
condensing
Pr
essure
465hpa to 1060 hPa
Elevation
altitude
-
378m to
60
00
m
(
-
1240 feet to
19685
feet)
Physical Characteristics
Dimensions (monitor enclosure)
W36XL
72
XH
24
mm
Weight
40
gram
Visual alarm
Low battery, system failur
e