EasyManua.ls Logo

Oxitone 1000M - Page 31

Oxitone 1000M
39 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Oxitone 1000M
User Manual Rev. 1.
2
Page 31 of 39
Display / Indicators
Data display: SpO
2
%, pulse rate (PR) beats per minute, battery level indicator, system status poor signal
indicator and skin temperature.
Type
OLED
Compliance
Equipment Classification
IEC 60601
-
1
Type of Protection (battery po
wer)
Internally powered
Accuracy Pulse Oximeter Equipment
ISO 80601
-
2
-
61
Degree of Protection
–
Sensor
Type BF
-
Applied Part
Mode of operation
Spot Check
Enclosure degree of ingress protection
IP 32
Biological evaluation
ISO10933
-
1:part 1
-
11
Usability
IEC60601
-
1
-
6
IEC60601
-
1
-
2:2014 (EMC)
Homecare IEC60601
-
1
-
11
Photobiological
safety of lamps
IEC 62471
ISO80601
-
2
-
56 Thermometers
Interfering Substances
Refer to Section 4.3. Safety Instructions
Note: This monitor complies with ISO 10993-1, Biological evaluation of medical
devices-part1.
Note: This product complies with RoHS.
Bluetooth Wireless Technology Information: Contains FCC ID: HSW2832
Bluetooth Compliance: Version 5.0 single mode low energy
Operating Frequency: 2.4 to 2.4835 GHz
Output Power: TX: +3 dBM
Operating Range: 10meter radius (line of sight)
Network Topology: Star - bus
Operation: Slave
Model 1000M
Anntena Type: Integrated chip type antenna
Modulation Type: Frequency Hopping spread Spectrum
Data Rate: 1Mbit/second
Data Latency: 6ms