Oxitone 1000M
User Manual Rev. 1.
2
Page 28 of 39
Condition
Possible Cause
Recommended Action
Battery does not charge
Device charging contacts (see
Sec' 5.1.2) contaminated
Clean contacts according to Sec' 8 cleaning
instructions
The APP cannot find the pulse
oximeter
The Bluetooth may not be
working
Reestablish the Bluetooth connection, if still
not successful, restart your PC.
Persistent system error
displayed. Power ON.
Hi voltage ESD
Power OFF the device and back
twice
.
If not working the device requires service
or replacement.
10.1. REPLACING THE STRAPS ON THE OXITONE 1000M
Follow the below steps to replace your Oxitone 1000M starps.
1. Slide the band’s spring bar inwards.
2. Pull the band away from the Oxitone 1000M body.