The device must be able to measure the pulse properly to obtain an accurate SPO2
measurement. Verify that nothing is hindering the pulse measurement before relying on the
SPO2 measurement.
Operation of this device below the minimum amplitude of 0.3% modulation may cause
inaccurate results.
General operation of the device may be affected by the use of an electrosurgical unit (ESU). This
device should not be used adjacent to other equipment. If adjacent use is necessary, the device
should be observed carefully to verify normal operation.
Keep the oximeter away from young children.
Oxitone 1000M monitor is intended for indoor operation.
The monitor should not be used as a substitute for laboratory blood analyzer.
Do not use before reading and understanding this user guide.
Not for use in shower, bath tub, sink or pool.
Excessive device wearing pressure for prolonged periods can induce pressure injury.
Only apply the device on clean, intact wrist skin.
Do not use the Oxitone 1000M outside the declared environmental conditions (see Section
Reference source not found.). Operating the device outside the declared environmental
conditions can lead to incorrect measurements.
Do not use the device when alarms are required.
Cautions
The device is designed to determine the percentage of arterial oxygen saturation of functional
hemoglobin. Factors that may degrade pulse oximeter performance or affect the accuracy
of the measurement include the following:
* Improperly applied device * Anemia or low hemoglobin
* Jewelry or ornaments on the tested hand wrist concentrations
* Poor pulse quality * Carboxyhemoglobin (COHB)
* Excessive motion * Venus pulsation
* Methemoglobin (Met Hb) * Dysfunctional hemoglobin
* Cardio green or indocyanine and other intravascular dyes
* Avoid Excessive light such as sunlight or direct home lighting and pulsating strobe lights
The device has no audible alarms and is intended for spot-checking.
The device may not work when circulation is reduced. Warm or rub the hand wrist area to
increase perfusion.
The device is designed to be attached to either hand wrist.