EasyManua.ls Logo

PerkinElmer Caliper IVIS Spectrum

PerkinElmer Caliper IVIS Spectrum
74 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
7. Operating the IVIS
®
Spectrum
40
2. Open the imaging chamber door to see the alignment grid projected by the
laser/galvanometer on the imaging platform
(Figure 7.5).
The field of view (FOV) setting defines the size of the squares in the
alignment grid.
3. To change the size of the squares in the alignment grid, select a different
FOV setting.
For example FOV settings, see Figure 6.4, page 28.
NOTICE
The alignment laser turns off after 60 seconds. Repeat step 1 to turn on the
laser.
4. Using the alignment grid as a guide, place the specimen(s) at the locations
of interest, then proceed with image acquisition.
NOTICE
The Living Image® software controls the photographic illumination LED
intensity.
Figure 7.4 IVIS System Control panel
Figure 7.5 Laser alignment grid
To display the alignment
grid, choose Alignment
Grid On.

Table of Contents

Related product manuals