Workstation software installation instructions
97
Installation
Environment
Normal clean laboratory conditions usually provide a satisfactory
operational environment but the following points should be taken into
consideration.
Ventilation should be adequate for all conditions of use
The temperature should be reasonably constant at about 22 ºC
Relative humidity should not be excessive
Direct sunlight should not be able to reach the instrument
If possible, there should be a separate power line for the instrument
itself with an isolation switch and fuse box.
If excessive fluctuations in the mains voltage are anticipated, a
mains stabilizer may be necessary.
Note! Make sure that when the instrument is at its place of operation it is
exactly level. This is important to ensure proper function of the stacker.
1. Unpacking instructions
Follow the instructions on the document attached on the instrument!
2. Installation of options
If the instrument has been ordered with stacker, dispenser or TRF LASER,
follow the instructions in the service manual or in the support folder on the
software CD!
If the Enhanced Security option is installed see Administrator Manual
Enhanced Security (supplied with the instrument) for User Name and
Password!
3. PC
Each instrument is supplied with a PC which has the CAN card, as well as
the needed software installed.
Note! The computer cannot be set to go automatically to a standby state.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the PerkinElmer EnVision and is the answer not in the manual?

PerkinElmer EnVision Specifications

General IconGeneral
Light SourceXenon flash lamp
SoftwareEnVision Manager
Read ModesAbsorbance, Fluorescence, Luminescence, Time-Resolved Fluorescence, Fluorescence Polarization
Plate Formats384-well
Detection TechnologyPMT
Temperature ControlAmbient to 45°C