Specifications
72
Measurement modes
• Fluorescence intensity from top
• Fluorescence intensity from top and bottom
• Time-resolved fluorescence from top
• Time-resolved fluorescence from top and bottom
• Multiple window TRF
• Homogenous TRF (LANCE, HTRF, LanthaScreen)
• Fluorescence polarization
• Absorbance
• Luminescence
• Ultra Sensitive Luminescence
• Dual emission
• AlphaScreen or HTS AlphaScreen
• Adjustable measurement height
• Dispensing
Plate shaking
Three plate shaking modes are available: linear, orbital and double orbital.
Three speed levels can be selected and the amplitude of the movement is
adjustable.
Scanning
Scanning of wells (several measuring points within a well) are available for
all technologies.
Kinetics
Repeated measurements or operations can be defined by plate or well.
Monochromator option
Two double monochromators are located inside the instrument casing. Light
fibers are used to guide the light to and from them. The excitation light from
the lamp is directed through the excitation double monochromator into the
sample. The emission light from the sample is directed through the emission
double monochromator to the detector.
Applications supported
• Photometric application range 230 – 1000 nm
• Photometric spectrum scan 230 – 1000 nm
• Fluorescence intensity application range 230 - 850 nm
• Fluorescence intensity excitation & emission spectrum scan
measurement 230 - 850 nm