Chapter 7
7-8 V60 Ventilator Service Manual 1049766 Rev A
Component Removal/Installation
7.7 Oxygen Inlet Follow these steps to remove the oxygen inlet (Figure 7-7). Reverse to install.
1. Disconnect all power (section 7.1).
2. Loosen the screws that attach the retaining bracket to the GDS, then
remove the bracket.
3. Pull to remove the oxygen inlet.
Figure 7-7: Removing the Oxygen Inlet
Oxygen
inlet
Retaining
bracket
M3 screw (x2)
(113 +
11.3 N cm/
10 +
1 in.-lbf)
Washer (x2)