Appendix A
1049766 Rev A V60 Ventilator Service Manual
A-13
17. Remove the null modem cable and adapter from the V60 ventilator.
18. Turn the V60 ventilator off by pressing and holding the On/Shutdown
button on the user interface.
19. Enter Diagnostic mode by pressing and holding the nav-ring Enter
button and turn on the ventilator by pressing the On/Shutdown button
on the user interface. Within 5 seconds, release and press the nav-ring
Enter button again to enter Diagnostic mode.
20. Touch Service and verify the Options on the V60 ventilator are the
same as the options listed on the GatewayInterface screen.
21. Click Exit on the GatewayInterface screen.
22. Turn the V60 ventilator off by pressing the On/Shutdown button.
23. Software reinitiation is complete.
A.4 Remote
Troubleshooting
Follow these steps to connect the V60 ventilator to Respi-Link and prepare for
remote troubleshooting.
NOTE: Do not connect the null modem cable until instructed to do so.
1. Turn on the PC and verify the Gateway Interface is running by double-
clicking on Start Auto Respi-Link Gateway in the Respi-Link folder on
your desktop.
2. If the ventilator is not already in Diagnostic mode, press and hold the
nav-ring Enter button and turn on the ventilator by pressing the
On/Shutdown button on the user interface. Within 5 seconds, press
the nav-ring Enter button again to enter Diagnostic mode.
3. Touch Service, then Pneumatics.
4. Connect a standard 9-pin male-female RS-232 null modem cable and
9-pin to 25-pin female-male adapter between the PC serial port and
the V60 ventilator.
5. The V60 ventilator is now connected to Respi-Link and can be
monitored remotely by a Respironics representative.
6. Contact your local service representative or Product Support at 800-
345-6443 for further instructions.