Chapter 7
Component Removal/Installation
1049766 Rev A V60 Ventilator Service Manual 7-33
7.30 Switch PCBA Follow these steps to remove the switch PCBA (Figure 7-31). Reverse to
install.
1. Open the UI (section 7.28).
2. Remove the screws that attach the switch PCBA to the inside of the
front bezel.
3. Disconnect the nav-ring and UI PCBA cables from the switch PCBA.
NOTE: During installation, verify that the nav-ring cable is fully seated into its
connector on the switch PCBA.
Figure 7-31: Removing the Switch PCBA
Switch PCBA
Front bezel
Switch PCBA to
UI PCBA cable
M3 x 6 screw (x4)
(62 +
5.7 N cm/
5.5 +
0.5 in.-lbf)
Nav-ring cable