EasyManua.ls Logo

Philips Respironics V60 - 6.1.1 Setting Up the Serial Interface Using HyperTerminal

Philips Respironics V60
246 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Chapter 6
6-2 V60 Ventilator Service Manual 1049766 Rev A
Reports and Software Downloads
6.1.1 Setting Up the Serial Interface Using HyperTerminal
Follow these steps to set up the HyperTerminal program, which enables the
ventilator to communicate to a service PC.
1. Launch the HyperTerminal program on the PC by clicking Start > All
Programs > Accessories > Communications > HyperTerminal.
2. If the Default Telnet Program? screen appears, click No (Figure 6-1).
Figure 6-1: Default Telnet Program Screen
3. Enter V60 as the name, choose an icon, then click OK (Figure 6-2).
Figure 6-2: Selecting the Connection Description

Table of Contents